Nitrogen washout/washin, helium dilution, and computed tomography in the assessment of end-expiratory lung volume

@inproceedings{Richard2013NitrogenWH,
  title={Nitrogen washout/washin, helium dilution, and computed tomography in the assessment of end-expiratory lung volume},
  author={J. C. Richard and Claude Gu{\'e}rin},
  booktitle={Critical care},
  year={2013}
}
We read thoroughly the article published by Chiumello and colleagues [1] in a past issue of Critical Care, along with its corresponding erratum [2], and we believe that the authors have erroneously computed the limits of agreement on several occasions in their article. Indeed, according to the original article by Bland and Altman [3], which was cited by the authors in their Materials and methods section, the limits of agreement should be computed as bias ± 2 standard deviations. Chiumello and… CONTINUE READING