A high performance magnetically-shielded room (MSR) has been built by Imedco, to house a 4-D MEG system, containing both gradiometers and magnetometers (Vectorview). The MSR consists of three nested enclosures, each consisting of a high-permeability magnetic layer plus an aluminum layer, where special care was given to the junctions of the inner aluminum layer. Measurements of the passive shielding factor yield 1,630 (64dB), 3,600 (71dB), 240,000 (107dB) , and 78,000,000 (158dB) at frequencies… 

Figures from this paper

Separable Magnetic Shield with Magnetic Shaking Enhancement
In this paper, we describe a separable cylindrical magnetic shield and a method for evaluating the magnetic shielding performance. A human size magnetic shield (Length: 1.5m, Inner diameter: O.5m)
Magnetic Field Stabilization for Magnetically Shielded Volumes by External Field Coils
Design and Optimization of Multilayer Cylindrical Magnetic Shield for SERF Atomic Magnetometer Application
TLDR
An optimization method to acquire a sufficiently effective, low-noise, and lightweight multi-layer shield using the nonlinear programming quadratic line (NLPQL) algorithm is presented.
Calibration and Thermal Analysis of a Conformal SERF Magnetometer Array for MEG Applications
In magnetoencephalography (MEG), a measurement of the magnetic field produced by the brain, it is critical for the distance between the magnetometers and the brain to be small and unchanging.
Passive Magnetic Shielding in Gradient Fields
The effect of passive magnetic shielding on dc magnetic field gradients imposed by both external and internal sources is studied. It is found that for concentric cylindrical or spherical shells of
An Approach to Analyzing Magnetically Shielded Room Comprising Discontinuous Layers Considering Permeability in Low Magnetic Field
To enable the realization of weak magnetic field at the level of several nanotesla, a magnetically shielded room (MSR) is required to create a space with residual magnetic field below 1 nT. Since the
Passive magnetic shielding in static gradient fields
The effect of passive magnetic shielding on dc magnetic field gradients imposed by both external and internal sources is studied for two idealized shield models: concentric spherical and
Development of a highly balanced gradiometer for fetal magnetocardiography
Activation of the cardiac muscle is associated with ion transport through the membrane of the cardiac cell. This gives rise to the strongest electrophysiological signals in human body: the
Dynamic Shielding in Biomagnetism
The paper presents an installation for biomagnetic studies and measurements in ambiental electromagnetic environment of the biomagnetic signals produced by the electrophysiological activities. The
Method development for ultra-low-field magnetic resonance imaging and magnetoencephalography
TLDR
Methods for MRI and MEG have been developed, with an emphasis on ultra-lowand low-field applications, and a new method for power correlations between brainwaves is developed and tested on MEG data.
...
...

References

SHOWING 1-2 OF 2 REFERENCES
Enhancement of Ferromagnetic Shielding against Low-Frequency Magnetic Fields
Experimental data are given to show that the impression of an alternating magnetic field of constant amplitude on a ferromagnetic shield, called shaking, enhances the shielding factor against
The 8-layered magnetically shielded room of the PTB
  • Biomag
  • 2000