NEMA NU 2-2001 performance measurements of an LYSO-based PET/CT system in 2D and 3D acquisition modes.

@article{Kemp2006NEMAN2,
  title={NEMA NU 2-2001 performance measurements of an LYSO-based PET/CT system in 2D and 3D acquisition modes.},
  author={Bradley J. Kemp and Chang Kim and John Joseph Williams and Alexander Ganin and Val J. Lowe},
  journal={Journal of nuclear medicine : official publication, Society of Nuclear Medicine},
  year={2006},
  volume={47 12},
  pages={1960-7}
}
UNLABELLED The National Electrical Manufacturers Association (NEMA) NU 2-2001 performance measurements were conducted on the Discovery RX, a whole-body PET/CT system under development by GE Healthcare. The PET scanner uses 4.2 x 6.3 x 30 mm lutetium yttrium orthosilicate (LYSO) crystals grouped in 9 x 6 blocks. There are 24 rings with 630 crystals per ring and the ring diameter is 88.6 cm. The transaxial and axial fields of view are 70.0 and 15.7 cm, respectively. The scanner has retractable… CONTINUE READING

Citations

Publications citing this paper.
Showing 1-10 of 42 extracted citations

References

Publications referenced by this paper.
Showing 1-10 of 19 references

Implementation and analysis of a 3D iterative reconstruction algorithm for the Discovery ST PET/CT scanners [abstract

  • M Iatrou, RM Manjeshwar, SG Ross
  • J Nucl Med
  • 2005
1 Excerpt

Similar Papers