Multisite evaluation of a continuous intraarterial blood gas monitoring system.

@article{Larson1994MultisiteEO,
  title={Multisite evaluation of a continuous intraarterial blood gas monitoring system.},
  author={Charles P Larson and Jeffery Vender and Adam Seiver},
  journal={Anesthesiology},
  year={1994},
  volume={81 3},
  pages={543-52}
}
BACKGROUND We compared the performance of a new, continuous intraarterial blood gas (CIABG) monitor with arterial values obtained periodically and analyzed by conventional equipment. METHODS A CIABG monitoring system consisting of a sterile, disposable, fiberoptic sensor and a microprocessor-controlled monitor with a self-contained calibration unit and detachable display panel was used. The sensor was inserted through a 20-G radial artery cannula. Light was transmitted from the monitor to the… CONTINUE READING