Monte Carlo study of the dependence of the KAP-meter calibration coefficient on beam aperture, x-ray tube voltage and reference plane.

Abstract

The Monte Carlo method was used to study the dependence of the calibration coefficient on the tube voltage, beam aperture and reference plane in simplified over-couch geometries modelling VacuTec's type 70157 KAP-meter both with and without an additional filter. The MCNP5 code was used to calculate (i) energy imparted to air cavities of the KAP-meter and… (More)

Topics

Figures and Tables

Sorry, we couldn't extract any figures or tables for this paper.

Slides referencing similar topics