Monte Carlo and experimental dosimetry of an I125 brachytherapy seed.

@article{Dolan2006MonteCA,
  title={Monte Carlo and experimental dosimetry of an I125 brachytherapy seed.},
  author={James Dolan and Zuofeng Li and Jeffrey F. Williamson},
  journal={Medical physics},
  year={2006},
  volume={33 12},
  pages={4675-4684}
}
We have performed a comprehensive dosimetric characterization of the Oncura™ model 6711 I125 seed using both experimental [LiF thermoluminscent dosimetry (TLD)] and theoretical (Monte Carlo photon transport) methods. In addition to determining the dosimetric parameters of the 6711, this report quantified: (1) the angular dependence of LiF TLD energy response functions for both point and volume detectors in water, poly(methylmethacrylate), and solid water media; and (2) the contribution of… CONTINUE READING
BETA

Citations

Publications citing this paper.
SHOWING 1-10 OF 14 CITATIONS

References

Publications referenced by this paper.
SHOWING 1-10 OF 33 REFERENCES

A comparison of solid phantoms with water for dosimetry of 125 I brachytherapy sources

  • J. A. Meli, R. Nath
  • Med . Phys .
  • 2003

New national airkermastrength standards for 125 I and 103 Pd brachytherapy seeds

  • P. J. Lamperti, R. Loevinger
  • J . Res . Natl . Inst . Stand . Technol .
  • 2003

Do - simetric characteristics , airkerma strength calibration and verification of Monte Carlo simulation for a new ytterbium169 brachytherapy source

  • J. F. Williamson, Z. Li, V. Mishra, A. S. Meigooni
  • Int . J . Radiat . Oncol . , Biol . , Phys .
  • 2002

Similar Papers