Measurementof Plasma Phenytoinby EMIT Ifl the Monarch CentrifugalAnalyzer : Studies on the Eliminationof Within-RotorDrift

@inproceedings{Wenham2004MeasurementofPP,
  title={Measurementof Plasma Phenytoinby EMIT Ifl the Monarch CentrifugalAnalyzer : Studies on the Eliminationof Within-RotorDrift},
  author={Wenham and Heather M. Barbour},
  year={2004}
}
We investigated the measurement of phenytoin in plasma by the EMIT#{174} method, in the Monarch centrifugal analyzer. When we used the standard protocol supplied by Instrumentation Laboratory, significant drift was observed across a full rotor loaded with 31 replicates of a single specimen. Although the cause remains obscure, the drift was eliminated by using a load-spin-reload-spin procedure. This makes it possible to analyze large batches of samples without excessive use of standards and… CONTINUE READING
0 Extracted Citations
9 Extracted References
Similar Papers

Referenced Papers

Publications referenced by this paper.