Measurement of plasma histamine by stable isotope dilution gas chromatography-mass spectrometry: methodology and normal values.

@article{Keyzer1984MeasurementOP,
  title={Measurement of plasma histamine by stable isotope dilution gas chromatography-mass spectrometry: methodology and normal values.},
  author={Jules J de Keyzer and Bert G. Wolthers and Frits A. J. Muskiet and Henk Breukelman and Henk F. Kauffman and Koert de Vries},
  journal={Analytical biochemistry},
  year={1984},
  volume={139 2},
  pages={474-81}
}
A new method for the determination of histamine by stable isotope dilution mass fragmentography is described. The method is specific, sensitive, and accurate, resulting in a within-day coefficient of variation of 4.1% and a day-to-day variation of 7.9%. It was shown that the first blood sample after a venipuncture can contain an artificially elevated plasma histamine concentration. Platelets contain about 7 pmol histamine/10(9) cells. Serum histamine was elevated about four times in comparison… CONTINUE READING