Measurement of intracompartmental pressure with use of a new electronic transducer-tipped catheter system.

@article{Willy1999MeasurementOI,
  title={Measurement of intracompartmental pressure with use of a new electronic transducer-tipped catheter system.},
  author={Christian Willy and Heinz Gerngross and J{\"u}rgen Sterk},
  journal={The Journal of bone and joint surgery. American volume},
  year={1999},
  volume={81 2},
  pages={158-68}
}
Laboratory and clinical tests were carried out to determine the clinical usefulness, validity, and safety of a new self-calibrating, battery-powered monitoring system for the measurement of intramuscular pressure with use of an electronic transducer-tipped catheter. The eight probes accurately recorded applied pressures ranging from zero to 160 millimeters of mercury (zero to 21.33 kilopascals). The system registered little temperature-induced drift (maximum, 1.25 millimeters of mercury [0.17… CONTINUE READING

Citations

Publications citing this paper.
Showing 1-10 of 23 extracted citations

The diagnosis of acute compartment syndrome: a review

European Journal of Trauma and Emergency Surgery • 2014
View 1 Excerpt

Similar Papers