MR Compatibility of Mechatronic Devices: Design Criteria

@inproceedings{Chinzei1999MRCO,
  title={MR Compatibility of Mechatronic Devices: Design Criteria},
  author={Kiyoyuki Chinzei and Ron Kikinis and Ferenc A. Jolesz},
  booktitle={MICCAI},
  year={1999}
}
Criteria to design MR (Magnetic Resonance) compatible mechatronic devices, e.g., surgical robots, are discussed. Some of critical interactions between MRI and mechatronic devices are discussed. Experimental results of the effects from several passive and active mechanical elements are demonstrated. Some passive elements, e.g., ball screw and linear guide have point contacts therefore, they are required to be made of hard smooth surfaces. Both beryllium-copper and a new stainless steel were… CONTINUE READING
Highly Cited
This paper has 387 citations. REVIEW CITATIONS
106 Citations
9 References
Similar Papers

Citations

Publications citing this paper.
Showing 1-10 of 106 extracted citations

388 Citations

02040'00'03'07'11'15
Citations per Year
Semantic Scholar estimates that this publication has 388 citations based on the available data.

See our FAQ for additional information.

References

Publications referenced by this paper.
Showing 1-9 of 9 references

MR Safety and MR Compatibility: Test Guidelines for Signa SP

  • GE Medical Systems ed
  • Version 1.0, http://www.ge.com/medical/mr/iomri…
  • 1997
Highly Influential
8 Excerpts

Pocket Guide to MR Procedures and Metallic Objects: Update 1998

  • F. G. Shellock
  • Lippincott-Raven publishers, Philadelphia
  • 1998
1 Excerpt

Piezoelectric ultrasonic motor - design and comparison

  • M. Crivii, M. Jufer
  • In proc Annu Symp Incremental Motion Control Syst…
  • 1994
1 Excerpt

Introduction to Ultrasonic Motor (Chouonpa Motor Nyuumon; written in Japanese)

  • T. Sashida, T. Kenjo
  • Sogo Denshi Shuppan, Tokyo,
  • 1991
1 Excerpt

Similar Papers