Initial stability and bone strain evaluation of the immediately loaded dental implant: an in vitro model study.

@article{Huang2011InitialSA,
  title={Initial stability and bone strain evaluation of the immediately loaded dental implant: an in vitro model study.},
  author={Heng-Li Huang and Y. Y. Chang and Dan-Jae Lin and Yu-fen Li and Kuan-ting Chen and Jui-Ting Hsu},
  journal={Clinical oral implants research},
  year={2011},
  volume={22 7},
  pages={
          691-698
        }
}
OBJECTIVES The aim of this study was to evaluate the effects of cortical bone thickness and trabecular bone elastic modulus on the strain in the bone surrounding an immediately loaded implant. We also examined the correlations between bone structure and the following indices of primary implant stability: insertion torque value (ITV), Periotest value (PTV), and implant stability quotient (ISQ). MATERIAL AND METHODS The ITV, PTV, and ISQ were measured in 24 artificial jaw bone models… CONTINUE READING
BETA

Citations

Publications citing this paper.
SHOWING 1-10 OF 23 CITATIONS

Timeframe of socket cortication after tooth extraction: A retrospective radiographic study.

  • Clinical oral implants research
  • 2018
VIEW 5 EXCERPTS
CITES BACKGROUND
HIGHLY INFLUENCED

Trabecular bone structural parameters evaluated using dental cone-beam computed tomography: cellular synthetic bones

  • Biomedical engineering online
  • 2013
VIEW 4 EXCERPTS
CITES METHODS & BACKGROUND
HIGHLY INFLUENCED

Variations in crestal cortical bone thickness at dental implant sites in different regions of the jawbone.

  • Clinical implant dentistry and related research
  • 2017
VIEW 1 EXCERPT
CITES BACKGROUND

Similar Papers

Loading similar papers…