Influence of the electrode compartment separating membranes on continuous flow electrophoresis.

The influence of the ion permeable diaphragms, separating the electrode compartments of continuous flow electrophoresis chambers from the separation space, are described in detail. A set of equations is derived from a simplified phenomenological model of the ion transport across media boundaries. These equations are useful to predict the magnitude of… CONTINUE READING