Influence of multiple use of Timedur-titanium cannon drills: thermal response and scanning electron microscopic findings.

@article{Jochum2000InfluenceOM,
  title={Influence of multiple use of Timedur-titanium cannon drills: thermal response and scanning electron microscopic findings.},
  author={R M Jochum and Peter A. Reichart},
  journal={Clinical oral implants research},
  year={2000},
  volume={11 2},
  pages={139-43}
}
Multiple use of dental drills is known to raise drilling temperatures. Therefore, to minimize surgical trauma, well sharpened drills are recommended. In order to evaluate surgical trauma due to multiple use, 20 Timedur cannon drills (ZL-Duraplant-Implant-System) were used 51 times in an in vitro experimental setting using pigs' mandibles. Preparations of implant sites were performed with cannon drills at a maximum rotational speed of 1200 r.p.m. with low pressure and water irrigation. Drills… CONTINUE READING