Inadvertent triggering of the ventilator caused by surgically placed pacer wires.

@article{France2012InadvertentTO,
  title={Inadvertent triggering of the ventilator caused by surgically placed pacer wires.},
  author={Timothy E France and A Randolph Garnett and David A Grooms},
  journal={Respiratory care},
  year={2012},
  volume={57 8},
  pages={1334-6}
}
Inadvertent ventilator triggering can occur for various reasons. Leaks in the ventilator circuit, endotracheal tube leaks, tracheal cuff leaks, cardiac oscillations, water condensate causing oscillations in the circuit tubing, ventilator expiratory valve integrity, and overly sensitive triggering mechanism settings may precipitate this phenomena. We present a case of inadvertent ventilator triggering caused by electrical stimulation of the diaphragm from surgically placed pacing wires post… CONTINUE READING