Inadvertent detection of 60-Hz alternating current by an implantable cardioverter defibrillator.

Abstract

A patient with an ICD received therapies from his ICD while exercising in an indoor swimming pool. Interrogation of the ICD revealed inappropriate detection of 60-Hz alternating current artifact and delivery of ICD therapies. 

Topics

Cite this paper

@article{Lee2002InadvertentDO, title={Inadvertent detection of 60-Hz alternating current by an implantable cardioverter defibrillator.}, author={Sung Woo Lee and Jeffrey P Moak and Brian Lewis}, journal={Pacing and clinical electrophysiology : PACE}, year={2002}, volume={25 4 Pt 1}, pages={518-9} }