In vivo feasibility of endovascular Doppler optical coherence tomography

@inproceedings{Sun2012InVF,
  title={In vivo feasibility of endovascular Doppler optical
coherence tomography},
  author={Cuiru Sun and Felix Nolte and Kyle H. Y. Cheng and Barry Vuong and Kenneth K. C. Lee and Beau A. Standish and Brian K. Courtney and Thomas R. Marotta and Adrian Mariampillai and Victor X. D. Yang},
  booktitle={Biomedical optics express},
  year={2012}
}
Feasibility of detecting intravascular flow using a catheter based endovascular optical coherence tomography (OCT) system is demonstrated in a porcine carotid model in vivo. The effects of A-line density, radial distance, signal-to-noise ratio, non-uniform rotational distortion (NURD), phase stability of the swept wavelength laser and interferometer system on Doppler shift detection limit were investigated in stationary and flow phantoms. Techniques for NURD induced phase shift artifact removal… CONTINUE READING

Citations

Publications citing this paper.
Showing 1-10 of 10 extracted citations

References

Publications referenced by this paper.
Showing 1-10 of 25 references

Feasibility of endovascular optical coherence tomography for high - resolution carotid vessel wall imaging

  • J. P. Cruz, T. R. Marotta, +5 authors B. A. Standish
  • Proc . SPIE
  • 2012

Similar Papers

Loading similar papers…