In vitro evaluation of aerosol bronchodilator delivery during noninvasive positive pressure ventilation: effect of ventilator settings and nebulizer position.

@article{Chatmongkolchart2002InVE,
  title={In vitro evaluation of aerosol bronchodilator delivery during noninvasive positive pressure ventilation: effect of ventilator settings and nebulizer position.},
  author={Sunisa Chatmongkolchart and Guilherme de Paula Pinto Schettino and Christine A Dillman and Robert M. Kacmarek and Dean R. Hess},
  journal={Critical care medicine},
  year={2002},
  volume={30 11},
  pages={2515-9}
}
OBJECTIVE Respiratory failure due to exacerbation of obstructive lung disease has been successfully treated with noninvasive positive pressure ventilation (NPPV). However, there have been no reports of factors affecting aerosol delivery during NPPV. Our objective was to determine the effect of ventilator settings and nebulizer position on albuterol delivery during NPPV. DESIGN Bench model study. SETTING University laboratory. SUBJECTS None. INTERVENTIONS A Respironics BiPAP S/T-D30 with… CONTINUE READING