Implementation of phantom-less IMRT delivery verification using Varian DynaLog files and R/V output.

Abstract

This study aims to evaluate the use of Varian radiotherapy dynamic treatment log (DynaLog) files to verify IMRT plan delivery as part of a routine quality assurance procedure. Delivery accuracy in terms of machine performance was quantified by multileaf collimator (MLC) position errors and fluence delivery accuracy for patients receiving intensity modulated… (More)
DOI: 10.1088/0031-9155/57/21/6761

Topics

Figures and Tables

Sorry, we couldn't extract any figures or tables for this paper.

Statistics

0102030201520162017
Citations per Year

Citation Velocity: 15

Averaging 15 citations per year over the last 3 years.

Learn more about how we calculate this metric in our FAQ.