ICD lead failure detection through high frequency impedance

Abstract

Abrasion-induced insulation breach is a common failure mode of silicone-body, transvenous, implantable cardioverter defibrillator leads. It is caused either by external compression or internal motion of conducting cables. The present method of monitoring lead integrity measures low frequency conductor impedance. It cannot detect insulation failures until… (More)
DOI: 10.1109/EMBC.2014.6945114

10 Figures and Tables

Cite this paper

@article{Kollmann2014ICDLF, title={ICD lead failure detection through high frequency impedance}, author={Daniel T. Kollmann and Charles D. Swerdlow and Mark W. Kroll and Gregory J. Seifert and Patrick A. Lichter}, journal={2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society}, year={2014}, pages={6487-6492} }