High-throughput proteomics using high-efficiency multiple-capillary liquid chromatography with on-line high-performance ESI FTICR mass spectrometry.

Abstract

We report on the design and application of a high-efficiency multiple-capillary liquid chromatography (LC) system for high-throughput proteome analysis. The multiple-capillary LC system using commercial LC pumps was operated at a pressure of 10,000 psi to deliver mobile phases through a novel passive feedback valve arrangement that permitted mobile-phase flow path switching and efficient sample introduction. The multiple-capillary LC system uses several serially connected dual-capillary column devices. The dual-capillary column approach eliminates the time delays for column regeneration (or equilibration) since one capillary column was used for a separation while the other was being washed. Several serially connected dual-capillary columns and electrospray ionization (ESI) sources were operated independently and can be used either for "backup" operation or for parallel operation with other mass spectrometers. This high-efficiency multiple-capillary LC system utilizes switching valves for all operations, enabling automated operation. The separation efficiency of the dual-capillary column arrangement, optimal capillary dimensions (column length and packed particle size), capillary regeneration conditions, and mobile-phase compositions and their compatibility with electrospray ionization were investigated. A high magnetic field (11.4 T) Fourier transform ion cyclotron resonance (FTICR) mass spectrometer was coupled on-line with this high-efficiency multiple-capillary LC system using an ESI interface. The capillary LC provided a peak capacity of approximately 650, and the 2-D capillary LC-FTICR analysis provided a combined resolving power of > 6 x 10(7) components. For yeast cytosolic tryptic digests > 100,000 polypeptides were detected, and approximately 1,000 proteins could be characterized from a single capillary LC-FTICR analysis using the high mass measurement accuracy (approximately 1 ppm) of FTICR, and likely more if LC retention time information were also exploited for peptide identification.

0204060'03'05'07'09'11'13'15'17
Citations per Year

196 Citations

Semantic Scholar estimates that this publication has 196 citations based on the available data.

See our FAQ for additional information.

Cite this paper

@article{Shen2001HighthroughputPU, title={High-throughput proteomics using high-efficiency multiple-capillary liquid chromatography with on-line high-performance ESI FTICR mass spectrometry.}, author={Ya-Ching Shen and Nikola Tolic and Rongbao Zhao and Ljiljana Pasa-Tolic and L Li and Scott J. Berger and Richard Harkewicz and Gordon A. Anderson and Mikhail E. Belov and Richard D. Smith}, journal={Analytical chemistry}, year={2001}, volume={73 13}, pages={3011-21} }