High-resolution functional vascular assessment with ultrasound

@article{Yeh2004HighresolutionFV,
  title={High-resolution functional vascular assessment with ultrasound},
  author={Chih-Kuang Yeh and Katherine W. Ferrara and Dustin E. Kruse},
  journal={IEEE Transactions on Medical Imaging},
  year={2004},
  volume={23},
  pages={1263-1275}
}
In order to improve the resolution of contrast-assisted imaging systems, we have created a high-frequency destruction/contrast replenishment imaging system with a spatial resolution of 160 /spl mu/m /spl times/ 160 /spl mu/m. The system utilizes a 1-MHz cylindrically focused transducer for destruction and a 25-MHz spherically focused transducer for pulse/echo imaging. Speckle tracking and a clutter filter are applied across frames to remove the challenging physiologic motion artifacts that are… CONTINUE READING