High performance liquid chromatography determination of cyanide in urine by pre-column fluorescence derivatization.

@article{Sano1989HighPL,
  title={High performance liquid chromatography determination of cyanide in urine by pre-column fluorescence derivatization.},
  author={Akira Sano and Mituo Takezawa and S Takitani},
  journal={Biomedical chromatography : BMC},
  year={1989},
  volume={3 5},
  pages={209-12}
}
A method for the determination of cyanide in human urine has been developed. The method is based on the reaction of cyanide with 2,3-naphthalenedialdehyde and taurine to give a fluorescent product for reversed-phase HPLC separation and fluorometric detection. After centrifugation followed by dilution of urine samples, the specimens could be analysed directly by this method. The recovery of cyanide added to urine at concentration levels of 50-1000 pmol/mL was 85-96%. The detection limit of… CONTINUE READING

From This Paper

Topics from this paper.
4 Citations
0 References