Hemodialyzer performance: a review of the trends over the past two decades.

@article{Woffindin1995HemodialyzerPA,
  title={Hemodialyzer performance: a review of the trends over the past two decades.},
  author={C Woffindin and Nicholas A. Hoenich},
  journal={Artificial organs},
  year={1995},
  volume={19 11},
  pages={1113-9}
}
Using data generated in the course of an evaluation program for hemodialyzers and associated devices, which was supported by the Medical Devices Agency of the Department of Health, we have compared the performance characteristics of 23 hemodialyzers evaluated 20 years ago with those of 18 hemodialyzers evaluated during 1993-1994. Surface areas of the two groups ranged from 0.6-2.5 m2 (mean 1.13 m2) and 0.6-1.8 m2 (mean 1.18 m2), respectively. The device performance has been compared both in… CONTINUE READING

From This Paper

Figures, tables, results, connections, and topics extracted from this paper.