GLC of cycloserine.

@article{Sondack1973GLCOC,
  title={GLC of cycloserine.},
  author={D L Sondack and Frank E. Gainer and H. J. Wesselman},
  journal={Journal of pharmaceutical sciences},
  year={1973},
  volume={62 8},
  pages={
          1344-6
        }
}
A GLC assay is described for cycloserine (D-4-amino-3-isoxazolidinone). The sample is dissolved in pyridine-water (2:1), an aliquot is evaporated, and the residue is treated with N,O-bis-(trimethylsilyl)acetamide and trimethylchlorosilane. After heating for a few minutes, chromatography is performed using a column containing methyl vinyl silicone, a relatively nonpolar liquid phase. Favorable quantitation is achieved by using hexamethylbenzene as the internal standard. The observed peak is the… CONTINUE READING