Filters for radiation reduction: a comparison.

@article{Kohn1988FiltersFR,
  title={Filters for radiation reduction: a comparison.},
  author={Melvin L. Kohn and A W Gooch and W. S. Keller},
  journal={Radiology},
  year={1988},
  volume={167 1},
  pages={
          255-7
        }
}
Aluminum, yttrium, copper, and R-filters were evaluated with respect to effective beam energy, entrance radiation exposure to the skin, x-ray tube loading, and the quality of resultant radiographs. Since all filters yielded acceptable radiographs, a figure of merit was developed to determine which filter delivered the lowest entrance radiation exposure for the lowest increase in x-ray tube loading. The figure of merit indicated that the best results were obtained with the 0.1-mm Cu filter with… CONTINUE READING