Fatigue resistance of engine-driven rotary nickel-titanium instruments produced by new manufacturing methods.

@article{Gambarini2008FatigueRO,
  title={Fatigue resistance of engine-driven rotary nickel-titanium instruments produced by new manufacturing methods.},
  author={Gianluca Gambarini and N M Grande and Gianluca Plotino and Francesco Somma and Manish Garala and Massimo De Luca and Luca Testarelli},
  journal={Journal of endodontics},
  year={2008},
  volume={34 8},
  pages={
          1003-5
        }
}
The aim of the present study was to investigate whether cyclic fatigue resistance is increased for nickel-titanium instruments manufactured by using new processes. This was evaluated by comparing instruments produced by using the twisted method (TF; SybronEndo, Orange, CA) and those using the M-wire alloy (GTX; Dentsply Tulsa-Dental Specialties, Tulsa, OK) with instruments produced by a traditional NiTi grinding process (K3, SybronEndo). Tests were performed with a specific cyclic fatigue… CONTINUE READING