False selection of syringe-brand compatibility and the method of correction during target-controlled infusion of propofol

@inproceedings{Chae2013FalseSO,
  title={False selection of syringe-brand compatibility and the method of correction during target-controlled infusion of propofol},
  author={Y J Chae and Jong-Yeop Kim and D W Kim and Bong-Ki Moon and S K Min},
  booktitle={Korean journal of anesthesiology},
  year={2013}
}
BACKGROUND We evaluated volumetric differences of syringe brand compatibilities, and investigated the impact of false brand settings on target-controlled infusion (TCI) and their methods of correction. METHODS Gravimetric measurement of 10 ml bolus infusions was performed using BD Plastipak (BDP) and Terumo compatible syringes, while setting to 7… CONTINUE READING