Failure of the MicroScan WalkAway system to detect heteroresistance to carbapenems in a patient with Enterobacter aerogenes bacteremia.

@article{Gordon2009FailureOT,
  title={Failure of the MicroScan WalkAway system to detect heteroresistance to carbapenems in a patient with Enterobacter aerogenes bacteremia.},
  author={N C Gordon and David William Wareham},
  journal={Journal of clinical microbiology},
  year={2009},
  volume={47 9},
  pages={3024-5}
}
We report the failure of the automated MicroScan WalkAway system to detect carbapenem heteroresistance in Enterobacter aerogenes. Carbapenem resistance has become an increasing concern in recent years, and robust surveillance is required to prevent dissemination of resistant strains. Reliance on automated systems may delay the detection of emerging resistance. 