Evolution of analytical performance in portable glucose meters in the last decade.

@article{Bhme2003EvolutionOA,
  title={Evolution of analytical performance in portable glucose meters in the last decade.},
  author={Philip B{\"o}hme and Mich{\`e}le Floriot and Marie-Aude Sirveaux and Danielle Durain and Olivier Ziegler and Pierre Drouin and Bruno Guerci},
  journal={Diabetes care},
  year={2003},
  volume={26 4},
  pages={
          1170-5
        }
}
OBJECTIVE To assess and compare the technical accuracy of portable glucose meters during the last decade. RESEARCH DESIGN AND METHODS One-thousand preprandial (pre) and postprandial (post) capillary whole-blood glucose values measured with meters owned mainly by diabetic patients were compared with a single laboratory method yearly from 1989 to 1999. A total of 21,950 capillary measurements and their corresponding laboratory reference values were analyzed at our clinic. RESULTS The lowest… CONTINUE READING

Figures, Tables, and Topics from this paper.

Citations

Publications citing this paper.
SHOWING 1-10 OF 16 CITATIONS

The accuracy of home glucose meters in hypoglycemia.

VIEW 1 EXCERPT
CITES RESULTS

References

Publications referenced by this paper.
SHOWING 1-10 OF 29 REFERENCES

Error detection and measurement in glucose monitors.

Hugkulstone CE: The effect of fluorescein on the accuracy of blood glucose determination by glucose meters

  • RJ Antcliff
  • Acta Ophthalmol Scand 77:197–199,
  • 1999
VIEW 1 EXCERPT

The effect of fluorescein on the accuracy of blood glucose determination by glucose meters

  • P Stenger, ME Allen, L Lisius
  • Acta Ophthalmol Scand
  • 1999

A 22,000 samples study 1174 DIABETES CARE, VOLUME 26, NUMBER 4, APRIL 2003

  • G Fanghanel, L Sanchez-Reyes, Morales
  • Arch Med Res
  • 1998
VIEW 1 EXCERPT