Evaluation of the mechanical behaviour of PathFile and ProGlider pathfinding nickel-titanium rotary instruments.

@article{Elnaghy2015EvaluationOT,
  title={Evaluation of the mechanical behaviour of PathFile and ProGlider pathfinding nickel-titanium rotary instruments.},
  author={A M Elnaghy and S E Elsaka},
  journal={International endodontic journal},
  year={2015},
  volume={48 9},
  pages={894-901}
}
AIM To assess and compare the resistance to cyclic fatigue, torsional stress, bending and buckling of ProGlider (PG; Dentsply Maillefer, Ballaigues, Switzerland) instruments with PathFile (PF; Dentsply Maillefer) pathfinding nickel-titanium rotary instruments. METHODOLOGY Size 16, .02 taper PG and PF instruments were rotated in simulated canals until failure, and the number of cycles to failure (NCF) was recorded to evaluate their cyclic fatigue resistance. Torsional strength was measured… CONTINUE READING