  title={Evaluation of the analytical performance of the Nova StatSensor creatinine meter and reagent strip technology for whole blood testing.},
  author={Kareena Leanne Schnabl and Sousan Bagherpoor and Pauline Diker and Christine Cursio and Jeffrey A. Dubois and Paul Yip},
  journal={Clinical biochemistry},
  year={2010},
  volume={43 12},
  pages={1026-9}
}
OBJECTIVE To evaluate the analytical performance of the Nova StatSensor creatinine meter. DESIGN AND METHODS Imprecision, linearity, patient correlation and interference studies were completed. RESULTS Total imprecision was 4.5-9.1%CV and a linear measurement range of 93-863 micromol/L (1.05-9.76 mg/dL) was verified. Whole blood creatinine results correlated well with laboratory plasma measurements (R(2)=0.9328) but exhibited a negative proportional bias. In vitro, high levels of creatine… CONTINUE READING