Evaluation of the Menarini-Arkray HA 8140 hemoglobin A1c analyzer.

@article{John1997EvaluationOT,
  title={Evaluation of the Menarini-Arkray HA 8140 hemoglobin A1c analyzer.},
  author={W G John and François Braconnier and Kor Miedema and Carlos Aulesa and Giovanna Piras},
  journal={Clinical chemistry},
  year={1997},
  volume={43 6 Pt 1},
  pages={968-75}
}
We describe a multinational evaluation of the Menarini-Arkray HA 8140 hemoglobin (Hb) A1c analyzer, which utilizes a high degree of automation, including bar code reading, cap piercing, and whole-blood sampling. With-in- and between-batch CVs were < 2%. Linearity was confirmed throughout the working range of the analyzer. Common Hb variants, including Hb S, Hb C, and Hb F, did not interfere with the Hb A1c separation, and the potentially interfering labile Schiff base was effectively removed… CONTINUE READING

Citations

Publications citing this paper.
Showing 1-10 of 12 extracted citations

References

Publications referenced by this paper.
Showing 1-10 of 11 references

Comparison of three commercially available ion-selective electrodes for ionized magnesium determination in serum: a two-center study [Abstract

  • HJ Huijgen, R Sanders, SA Cecco, NN Rehak, GT Sanders, RJ. Elin
  • Clin Chem
  • 1998
1 Excerpt

Comparison of two AVL analyzers for ionized magnesium determination [Abstract

  • SA Cecco, HJ Huijgen, R Sanders, NN Rehak, GTB Sanders, RJ. Elin
  • Clin Chem 1998;44:A114
  • 1998
1 Excerpt

A method for the measurement of ionized magnesium

  • D Filos, AO Okorodudu
  • Lab Med
  • 1995
2 Excerpts

Similar Papers