Evaluation of the Automated Haematology Analyser Sysmex M-2000

@inproceedings{Pohland1989EvaluationOT,
  title={Evaluation of the Automated Haematology Analyser Sysmex M-2000},
  author={D. Pohland},
  booktitle={Journal of clinical chemistry and clinical biochemistry. Zeitschrift fur klinische Chemie und klinische Biochemie},
  year={1989}
}
  • D. Pohland
  • Published in
    Journal of clinical chemistry…
    1989
  • Medicine
The automated Sysmex M-2000 was evaluated according to the ICSH (International Committee for Standardization in Haematology) protocol. After dilution of packed cells with cell-free plasma, blood cell counts were linear. The overall precision of the measured parameters was good; the CV's ranged between 0.64% and 2.06%. The carry-over was negligible; platelets showed the biggest carry-over with 0.25% in the Whole Blood Mode, while red blood cells (RBC) showed a carry-over of 0.55% in the… Expand
Automated Blood Count Analysis by Trimodal Size Distribution of Leukocytes with the SYSMEX E-5000
  • W. Bürgi, H. Marti
  • Medicine
  • Journal of clinical chemistry and clinical biochemistry. Zeitschrift fur klinische Chemie und klinische Biochemie
  • 1989
TLDR
In the present study, 31% of the microscopic blood cell differential counts were saved by using the SYSMEX E-5000, and the predictive values were also high for both pathological and normal results. Expand
Evaluation of a novel haematology analyser for use with feline blood.
TLDR
The haematology analyser was found to be highly reliable for the analysis of feline blood samples in a large veterinary laboratory and demonstrated very low imprecision. Expand
Evaluation of the LaserCyte: an in-house hematology analyzer for dogs and cats
TLDR
It was concluded that the LaserCyte allowed reliable determination of the RBC parameters, WBCs, neutrophils in both species and platelets in dogs and cats and is considered a useful in-house analyzer for the veterinary practice. Expand
Comparison between manual procedure and automated for determinant of WBCs and PCV in maternity and labor hospital in Karbala city
TLDR
It can be concluded that the automated analyzer (WBCs) and (Hb) were well correlated with manual standard method for two hematologic parameter so health care workers in the laboratory can save performance time for hematological parameter in automated analyzers. Expand
Accuracy of point‐of‐care‐testing (POCT) for determining hemoglobin concentrations
TLDR
The aim of the present study was to evaluate the accuracy and precision of point‐of‐care testing for blood hemoglobin concentration (cHb in g/l) measurements in critically ill patients. Expand
Diagnostic performance of automated Analyzers versus manual method for WBCs measurement among Population of Wad Madani, Gezira State, Sudan
TLDR
Two new approaches are proposed that combine radiolysis and EMT, which combine with each other to provide real-time information about the EMT treatment of central giant cell granuloma. Expand
Evaluierung des Hämatologieautomaten Sysmex SF-3000 unter besonderer Berücksichtigung pädiatrischer und onkologischer Aspekte - Evaluation of the Hematology Analyzer Sysmex SF-3000 with Special Regard to Pediatric and Oncologic Aspects
TLDR
Die Beurteilung der klinischen Eignung der durchflußzytometrischen Leukozytendifferenzierung erfolgte durch Vergleich mit dem 400-ZellenDifferentialblutbild bei 375 Blutprobenproben, von denen 223 bei der mikroskopischen Untersuchung pathologische Befunde aufwiesen. Expand
Cómo obtener un extendido de sangre periférica de óptima calidad
Resumen: a pesar de los grandes avances tecnologicos y el desarrollo de nuevos parametros en hematologia, el extendido de sangre periferica continua siendo la “prueba reina” en el diagnostico de lasExpand

References

SHOWING 1-2 OF 2 REFERENCES
Angewandte Statistik, 6th Edit., SpringerVerlag, Berlin-Heidelberg-New York-Tokio
  • 1984
Standard for Spedmen Collection , Part 1 : Blood Containers
  • 1983