Evaluation of techniques for the elimination of "hot" bladder artifacts in SPECT of the pelvis.

@article{OConnor1990EvaluationOT,
  title={Evaluation of techniques for the elimination of "hot" bladder artifacts in SPECT of the pelvis.},
  author={M Kevin O'Connor and Brian J. Kelly},
  journal={Journal of nuclear medicine : official publication, Society of Nuclear Medicine},
  year={1990},
  volume={31 11},
  pages={1872-5}
}
The purpose of this study was to validate the usefulness of two digital filtering techniques used to eliminate the artifacts caused by rapid bladder filling during SPECT of the pelvis. A dynamic phantom model was used containing two hips and a bladder. The phantom was studied under three conditions--bladder empty, filling, and full. The ability of the pixel truncation and interpolative background replacement techniques to eliminate bladder artifacts was assessed. Both techniques gave similar… CONTINUE READING

Citations

Publications citing this paper.
Showing 1-10 of 11 extracted citations

Reducing bladder artifacts in clinical pelvic SPECT images.

Journal of nuclear medicine : official publication, Society of Nuclear Medicine • 2004

Nonosseous abnormalities on bone scans.

Journal of nuclear medicine technology • 2003
View 1 Excerpt

Similar Papers