Evaluation of inspiratory rise time and inspiration termination criteria in new-generation mechanical ventilators: a lung model study.

@article{Chatmongkolchart2001EvaluationOI,
  title={Evaluation of inspiratory rise time and inspiration termination criteria in new-generation mechanical ventilators: a lung model study.},
  author={Sunisa Chatmongkolchart and Purris F Williams and Dean R. Hess and Robert M. Kacmarek},
  journal={Respiratory care},
  year={2001},
  volume={46 7},
  pages={666-77}
}
INTRODUCTION Inspiratory rise time adjustment during pressure ventilation and inspiration termination criteria adjustment during pressure support ventilation are available on some of the newest mechanical ventilators. Both are designed to improve patient-ventilator synchrony. However, the function of these adjuncts during pressure ventilation on these ventilators has not been evaluated. METHODS Three inspiratory rise times (minimum, medium, and maximum) were evaluated in 5 new-generation… CONTINUE READING

Citations

Publications citing this paper.
Showing 1-10 of 15 extracted citations

References

Publications referenced by this paper.
Showing 1-10 of 20 references

An analysis of desynchronization between the spontaneously breathing patient and ventilator during inspiratory pressure support

  • G Bonmarchand, V Chevron, +4 authors J. Leroy
  • Chest
  • 1995