Evaluation of ease of intubation with the GlideScope or Macintosh laryngoscope by anaesthetists in simulated easy and difficult laryngoscopy.

@article{Lim2005EvaluationOE,
  title={Evaluation of ease of intubation with the GlideScope or Macintosh laryngoscope by anaesthetists in simulated easy and difficult laryngoscopy.},
  author={Tsyr Jong Lim and Yi Jin Genevieve Lim and Eugene Hern Choon Liu},
  journal={Anaesthesia},
  year={2005},
  volume={60 2},
  pages={
          180-3
        }
}
The GlideScope is a new video laryngoscope developed for management of the difficult airway. We compared the GlideScope with the Macintosh laryngoscope in simulated easy and difficult laryngoscopy. Twenty anaesthetists were allowed three attempts to intubate in each of four laryngoscopy scenarios in a high fidelity simulator. In the simulated easy laryngoscopy scenarios, the anaesthetists took longer to intubate using the GlideScope than the Macintosh laryngoscope (mean (SD) 19.0 (9.7) s vs. 12… CONTINUE READING
BETA
Tweets
This paper has been referenced on Twitter 1 time. VIEW TWEETS

From This Paper

Topics from this paper.

Citations

Publications citing this paper.
SHOWING 1-10 OF 60 CITATIONS, ESTIMATED 34% COVERAGE

180 Citations

01020'69'85'02'19
Citations per Year
Semantic Scholar estimates that this publication has 180 citations based on the available data.