Errors in 14 pulse oximeters during profound hypoxia

@article{Severinghaus1989ErrorsI1,
  title={Errors in 14 pulse oximeters during profound hypoxia},
  author={John Wendell Severinghaus and Karen H. Naifeh and Shin Ok Koh},
  journal={Journal of Clinical Monitoring},
  year={1989},
  volume={5},
  pages={72-81}
}
The accuracy of pulse oximeters from fourteen manufacturers was tested during profound brief hypoxic plateaus in 125 subject sets using 50 normal adult volunteers, of whom 29 were studied two to nine times. A data set usually consisted of 10 subjects, and 13 sets were collected between August 1987 and July 1988. In the first 6 sets, six 30-second hypoxic plateaus were obtained per subject at 55±6% oxyhemoglobin (O2Hb) (range, 40 to 70%). In the last 7 sets, three hypoxic plateaus were obtained… CONTINUE READING
BETA

Citations

Publications citing this paper.
SHOWING 1-10 OF 60 CITATIONS, ESTIMATED 22% COVERAGE

FILTER CITATIONS BY YEAR

1990
2018

CITATION STATISTICS

  • 1 Highly Influenced Citations

  • Averaged 2 Citations per year over the last 3 years

References

Publications referenced by this paper.
SHOWING 1-9 OF 9 REFERENCES

Similar Papers