Enhanced quality factors and force sensitivity by attaching magnetic beads to cantilevers for atomic force microscopy in liquid

@article{Hoof2012EnhancedQF,
  title={Enhanced quality factors and force sensitivity by attaching magnetic beads to cantilevers for atomic force microscopy in liquid},
  author={Sebastian Hoof and Nitya Nand Gosvami and Bart W. Hoogenboom},
  journal={Journal of Applied Physics},
  year={2012},
  volume={112},
  pages={114324}
}
Dynamic-mode atomic force microscopy (AFM) in liquid remains complicated due to the strong viscous damping of the cantilever resonance. Here, we show that a high-quality resonance (Q>20) can be achieved in aqueous solution by attaching a microgram-bead at the end of the nanogram-cantilever. The resulting increase in cantilever mass causes the resonance frequency to drop significantly. However, the force sensitivity—as expressed via the minimum detectable force gradient—is hardly affected… 

Figures and Tables from this paper

Microinductor-Fused Atomic Force Microscopy Cantilevers for Dynamic Imaging and Multimodal Manipulation

New AFM cantilevers fabricated with a planar microcoil at the tip region, which can be used to generate or sense highly-localized magnetic fields and generated localized fields to selectively stimulate and monitor structural changes in viable cells with nm-scale detail are demonstrated.

Simultaneously Measuring Force and Displacement: Calibration of Magnetic Torque Actuated Microcantilevers for Nanomechanical Mapping

Force-distance curve-based atomic force microscopy (FD-based AFM) has been extensively used to study the interactive phenomena and mechanical properties of various materials at the nanoscale. Unlike

An Atomic Force Microscope with Dual Actuation Capability for Biomolecular Experiments

A modular atomic force microscope design for biomolecular experiments that can be operated in different modes, such as molecular pulling or force-clamp, by actuating the cantilever with the available actuators and has demonstrated in unbinding experiments of the biotin-streptavidin complex.

Nano-rheology of hydrogels using direct drive force modulation atomic force microscopy.

A magnetic force-based direct drive modulation method to measure local nano-rheological properties of soft materials across a broad frequency range using colloid-attached atomic force microscope probes in liquid highlights possible transitions in the probe-material contact mechanical behaviour for soft matter.

Direct torsional actuation of microcantilevers using magnetic excitation

Torsional mode dynamic force microscopy can be used for a wide range of studies including mapping lateral contact stiffness, torsional frequency or amplitude modulation imaging, and dynamic friction

Stochastic resonance induced performance enhancement of MEMS cantilever biosensors

This paper presents an application of stochastic resonance for improving the performance of micro-electromechanical system (MEMS) cantilever biosensors operating in liquid media. The hydrodynamic

Effective quality factor tuning mechanisms in micromechanical resonators

Quality factor (Q) is an important property of micro- and nano-electromechanical (MEM/NEM) resonators that underlie timing references, frequency sources, atomic force microscopes, gyroscopes, and

Efficiency improvement in the cantilever photothermal excitation method using a photothermal conversion layer

The proposed method, using a PTC layer made of colloidal graphite, greatly enhances photothermal excitation efficiency even for a relatively stiff cantilever in liquid.

Theoretical Analysis of Laterally Vibrating Hammerhead Microcantilever Sensors in a Viscous Liquid

THEORETICAL ANALYSIS OF LATERALLY VIBRATING HAMMERHEAD MICROCANTILEVER SENSORS IN A VISCOUS LIQUID Jinjin Zhang, B.E., M.E. Marquette University, 2013 Dynamically driven prismatic microcantilevers

References

SHOWING 1-10 OF 39 REFERENCES

Frequency noise in frequency modulation atomic force microscopy.

The contribution of the displacement Sensor noise to the frequency noise in the FM-AFM is described in detail to show how it is important to reduce the noise-equivalent displacement in the displacement sensor especially in low-Q environments.

High Q optical fiber tips for NC-AFM in liquid

This paper presents a tip based on an optical fiber which can operate in liquid with Q factors in excess of 100 using a 'diving bell' arrangement which allows only a small portion of the tip to be submerged and demonstrates stable imaging and force spectroscopy using this set-up.

Piconewton regime dynamic force microscopy in liquid

This work uses a mixed signal composed of a standard driving signal and a feedback signal consisting of the amplified and phase shifted oscillation signal to identify the resonance frequency of the cantilever by mechanical excitation of the fluid.

Coupling of conservative and dissipative forces in frequency-modulation atomic force microscopy

Frequency modulation atomic force microscopy FM-AFM utilizes the principle of self-excitation to ensure the cantilever probe vibrates at its resonant frequency, regardless of the tip-sample

Studies of vibrating atomic force microscope cantilevers in liquid

An atomic force microscope (AFM) design providing a focused spot of order 7 μm in diameter was used to analyze the motion of vibrating cantilevers in liquid. Picking an operating frequency for

Magnetic and acoustic tapping mode microscopy of liquid phase phospholipid bilayers and DNA molecules

We have constructed a fluid cell for an atomic force microscope that operates in tapping mode using either an oscillating piezo or magnetic drive. This fluid cell allows direct comparison of the

Atomic force microscope with magnetic force modulation

We have constructed a scanned stylus atomic force microscope (AFM) with direct force modulation and integrated microfluorescence optics. The instrument was designed to image the surface of massive

Frequency response of cantilever beams immersed in viscous fluids with applications to the atomic force microscope: Arbitrary mode order

The frequency response of a cantilever beam is well known to depend strongly on the fluid in which it is immersed. In this article, we present a theoretical model for the frequency response of a

Potential of interferometric cantilever detection and its application for SFM/AFM in liquids

An optical cantilever deflection detector with a spot size <3 µm and fm Hz(-1/2) sensitivity over a>10 MHz bandwidth is developed, enabling it to reach the thermal noise limit in scanning or atomic force microscopy experiments with stiff cantilevers.