Empirical formulae for calculating γ-ray detectors effective solid angle ratios
@inproceedings{Khatib2017EmpiricalFF, title={Empirical formulae for calculating $\gamma$-ray detectors effective solid angle ratios}, author={Ahmed M. El Khatib and Mohamed S. Badawi and Mohamed Abd Elzaher and Mona M. Gouda and A. A. Thabet and Mahmoud I. Abbas and Kholud. S. Almugren}, year={2017} }
Determination of the detector effi ciency using volumetric cylindrical sources is very important in various scientifi c and industrial fi elds, especially in the fi eld of quantitative analysis. To calculate the absolute activity of any sample, the full-energy peak effi ciency (FEPE) of the detector is needed. By applying the effi ciency transfer method, the FEPE of the detector would be determined easily without using the standard sources. This approach depends on two main factors. The fi rst…
Figures and Tables from this paper
References
SHOWING 1-10 OF 19 REFERENCES
Calculation of the Peak Efficiency for NaI(Tl) Gamma Ray Detector Using the Effective Solid Angle Method
- Physics
- 2012
In this work the Full-Energy Peak Efficiency (FEPE) of NaI(Tl) - scintillation detectors (5.08x5.08 cm 2 and 7.62x7.62 cm 2 ) values are calculated for coaxial cylindrical sources with radii greater…
An empirical formula to calculate the full energy peak efficiency of scintillation detectors.
- PhysicsApplied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine
- 2013
Calibration of cylindrical detectors using a simplified theoretical approach.
- PhysicsApplied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine
- 2006
Intercomparison of efficiency transfer software for gamma-ray spectrometry.
- Environmental ScienceApplied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine
- 2001
HPGE DETECTOR ABSOLUTE FULL-ENERGY PEAK EFFICIENCY CALIBRATION INCLUDING COINCIDENCE CORRECTION FOR CIRCULAR DISC SOURCES
- Engineering
- 2006
A direct mathematical approach is presented to calculate the absolute full-energy peak (photopeak) efficiency and coincidence correction of hyper pure germanium (HPGe) detectors for extended circular…
Coincidence summing corrections in Ge(Li)-spectrometry at low source-to-detector distances
- Environmental Science
- 1979
Analytical formulae for well-type NaI (Tl) and HPGe detectors efficiency computation.
- EngineeringApplied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine
- 2001
Calculation of relative full-energy peak efficiencies of well-type detectors
- Engineering, Physics
- 2002