Electrically compensated Fourier transform ion cyclotron resonance cell for complex mixture mass analysis.

@article{Kaiser2011ElectricallyCF,
  title={Electrically compensated Fourier transform ion cyclotron resonance cell for complex mixture mass analysis.},
  author={Nathan K. Kaiser and Joshua J Savory and Amy M McKenna and J. M. P. Quinn and Christopher L. Hendrickson and Alan G. Marshall},
  journal={Analytical chemistry},
  year={2011},
  volume={83 17},
  pages={6907-10}
}
Complex natural organic mixtures such as petroleum require ultrahigh mass spectral resolution to separate and identify thousands of elemental compositions. Here, we incorporate a custom-built, voltage-compensated ICR cell for Fourier transform ion cyclotron resonance mass spectrometry (FTICR MS), based on a prior design by Tolmachev to produce optimal mass resolution. The compensated ICR cell installed in a custom-built 9.4 T FTICR mass spectrometer consists of seven cylindrical segments with… CONTINUE READING