Effects of fabrication, finishing, and polishing procedures on preload in prostheses using conventional "gold' and plastic cylinders.

@article{Carr1996EffectsOF,
  title={Effects of fabrication, finishing, and polishing procedures on preload in prostheses using conventional "gold' and plastic cylinders.},
  author={Alan B. Carr and John B Brunski and Eileen Hurley},
  journal={The International journal of oral & maxillofacial implants},
  year={1996},
  volume={11 5},
  pages={589-98}
}
This study reviews fundamental concepts related to the use of screws and presents data describing the effect of fabrication, finishing, and polishing procedures on as-received preload for implant cylinders. Specifically, this study measured and compared preload produced when using as-received gold cylinders (the reference or gold standard), and cast cylinders produced from premade gold and plastic cylinders in the as-cast condition and following postcast finishing and polishing manipulations… CONTINUE READING