Effective doses from cone beam CT investigation of the jaws.

@article{Davies2012EffectiveDF,
  title={Effective doses from cone beam CT investigation of the jaws.},
  author={John Graham Davies and Ben Johnson and Na Drage},
  journal={Dento maxillo facial radiology},
  year={2012},
  volume={41 1},
  pages={30-6}
}
OBJECTIVES The purpose of the study was to calculate the effective dose delivered to the patient undergoing cone beam (CB) CT of the jaws and maxillofacial complex using the i-CAT Next Generation CBCT scanner (Imaging Sciences International, Hatfield, PA). METHODS A RANDO® phantom (The Phantom Laboratory, Salem, NY) containing thermoluminence dosemeters were scanned 10 times for each of the 6 imaging protocols. Effective doses for each protocol were calculated using the 1990 and approved 2007… CONTINUE READING