EPID dosimetry for pretreatment quality assurance with two commercial systems

@inproceedings{Bailey2012EPIDDF,
  title={EPID dosimetry for pretreatment quality assurance with two commercial systems},
  author={Daniel W. Bailey and Lalith K. Kumaraswamy and Mohammad Hashem Bakhtiari and Harish Kumar Malhotra and M. Podgorsak},
  booktitle={Journal of applied clinical medical physics},
  year={2012}
}
This study compares the EPID dosimetry algorithms of two commercial systems for pretreatment QA, and analyzes dosimetric measurements made with each system alongside the results obtained with a standard diode array. 126 IMRT fields are examined with both EPID dosimetry systems (EPIDose by Sun Nuclear Corporation, Melbourne FL, and Portal Dosimetry by Varian Medical Systems, Palo Alto CA) and the diode array, MapCHECK (also by Sun Nuclear Corporation). Twenty-six VMAT arcs of varying modulation… CONTINUE READING
Highly Cited
This paper has 38 citations. REVIEW CITATIONS

Citations

Publications citing this paper.
Showing 1-10 of 19 extracted citations

References

Publications referenced by this paper.
Showing 1-10 of 42 references

Comparison of the efficacy of MapCHECK and portal dosimetry in the pretreatment QA of IMRT treatment plans [abstract

  • J Liu, Z Wang, L Kumaraswamy, D Nazareth, M Podgorsak, H. Malhotra
  • Med Phys. ;35(6):2764
  • 2008
Highly Influential
7 Excerpts