Dynamic performance of accommodating intraocular lenses in a negative feedback control system: A simulation-based study

@article{Schor2007DynamicPO,
  title={Dynamic performance of accommodating intraocular lenses in a negative feedback control system: A simulation-based study},
  author={Clifton M. Schor and Shrikant R. Bharadwaj and Christopher D. Burns},
  journal={Computers in biology and medicine},
  year={2007},
  volume={37 7},
  pages={1020-35}
}
A dynamic model of ocular accommodation is used to simulate the stability and dynamic performance of accommodating intraocular lenses (A-IOLs) that replace the hardened natural ocular lens that is unable to change focus. Accommodation simulations of an older eye with A-IOL materials having biomechanical properties of a younger eye illustrate overshoots and oscillations resulting from decreased visco-elasticity of the A-IOL. Stable dynamics of an A-IOL are restored by adaptation of phasic and… CONTINUE READING

References

Publications referenced by this paper.
Showing 1-10 of 64 references

W

  • H. A. Weeber, G. Eckert, F. Soergel, C. H. Meyer
  • Pechhold, R.G.L. van der Heijde, Dynamic…
  • 2005
Highly Influential
9 Excerpts

Age-related changes in the accommodation mechanism, Optom

  • A.P.A. Beers, G. L. van der Heijde
  • Vis. Sci
  • 1996
Highly Influential
6 Excerpts

Graebel, Elasticity of tissues involved in accommodation

  • W.P.G.W.H.M. van Alphen
  • Vis. Res
  • 1991
Highly Influential
10 Excerpts

Similar Papers