Dose Optimization in TOF-PET/MR Compared to TOF-PET/CT

@inproceedings{Queiroz2015DoseOI,
  title={Dose Optimization in TOF-PET/MR Compared to TOF-PET/CT},
  author={Marcelo A Queiroz and Gaspar Delso and Scott D. Wollenweber and Timothy Deller and Konstantinos G. Zeimpekis and Martin W. H{\"u}llner and Felipe de Galiza Barbosa and Gustav K von Schulthess and Patrick Veit-Haibach and Kewei Chen},
  booktitle={PloS one},
  year={2015}
}
PURPOSE To evaluate the possible activity reduction in FDG-imaging in a Time-of-Flight (TOF) PET/MR, based on cross-evaluation of patient-based NECR (noise equivalent count rate) measurements in PET/CT, cross referencing with phantom-based NECR curves as well as initial evaluation of TOF-PET/MR with reduced activity. MATERIALS AND METHODS A total of 75 consecutive patients were evaluated in this study. PET/CT imaging was performed on a PET/CT (time-of-flight (TOF) Discovery D 690 PET/CT… CONTINUE READING
Recent Discussions
This paper has been referenced on Twitter 4 times over the past 90 days. VIEW TWEETS

Citations

Publications citing this paper.
Showing 1-10 of 10 extracted citations

References

Publications referenced by this paper.
Showing 1-10 of 30 references

PET perfor- mance measurements using the NEMANU 2–2001 standard

  • ME Daube-Witherspoon, JS Karp, ME Casey, FP DiFilippo, H Hines, G Muehllehner
  • J Nucl Med
  • 2001