Dilatation of Stent-Grafts by Luminal Pressures: Experimental Evaluation of Polytetrafluoroethylene (PTFE) and Woven Polyester Grafts

@article{Suzuki2001DilatationOS,
  title={Dilatation of Stent-Grafts by Luminal Pressures: Experimental Evaluation of Polytetrafluoroethylene (PTFE) and Woven Polyester Grafts},
  author={Kojiro Suzuki and Tsuneo Ishiguchi and Shoji Kawatsu and Hiroyoshi Iwai and Kunihiro Maruyama and Takeo Ishigaki},
  journal={CardioVascular and Interventional Radiology},
  year={2001},
  volume={24},
  pages={94-98}
}
To evaluate the effect on dilatation of three types of covering materials used for endovascular stent-grafts. Stent-grafts with three types of covering material [0.1-mm polytetrafluoroethylene (PTFE), 0.2-mm PTFE, and 0.15-mm woven polyester] were placed in a fluid circuit. For the pulsatile pressure test, the luminal pressure of 190/130 mmHg was loaded up to 300,000 pulses. For the static pressure test, the luminal pressure was increased from 50 to 300 mmHg at 50-mmHg increments. The percent… CONTINUE READING