Development and validation of an LC-ESI-MS method for quantitative determination of aripiprazole in human plasma and an application to pharmacokinetic study.

@article{Ravinder2012DevelopmentAV,
  title={Development and validation of an LC-ESI-MS method for quantitative determination of aripiprazole in human plasma and an application to pharmacokinetic study.},
  author={Sreedasyam Ravinder and Akula Tukaram Bapuji and Khagga Mukkanti and Datla Rama Raju and Hamsa Laxmi Venkata Ravikiran and Dendhi Chandrapal Reddy},
  journal={Journal of chromatographic science},
  year={2012},
  volume={50 10},
  pages={893-901}
}
A selective, sensitive, high pressure liquid chromatography-positive electrospray ionization tandem mass spectrometry method was developed and validated for the quantification of aripiprazole in human K(2)EDTA plasma using zolpidem tartrate as an internal standard. The analyte and internal standard were extracted from human plasma by solid-phase extraction using methanol. The eluted samples were chromatographed on a Grace Smart RP 18 4.6 × 100 mm, 3 µ column by using a 95:5 v/v mixture of… CONTINUE READING
3 Citations
14 References
Similar Papers

References

Publications referenced by this paper.
Showing 1-10 of 14 references

Development and validation of a high-performance liquid chromatography method using diode array detection for the simultaneous quantification of aripiprazole dehydro aripiprazole in human plasma

  • L. Frederique, D. Kayssa, T. Khalid, K. Linda, B. Sophie, P Pascal
  • Journal of Chromatography B,
  • 2008

LC-ESI-MS for rapid and sensitive determination of aripiprazole in human plasma

  • C. Z. Xiao, F. Wang, P. Xu, R. Zhu, H. Li
  • Chromatographia,
  • 2006

Development and validation of an LC-MS/MS method for the quantitative determination of aripiprazole and its main metabolite, OPC-14857, in human plasma

  • K. Masanori, Y. Mizooku, Y. Hirao, T. Osumi
  • Journal of Chromatography B,
  • 2005