Determination of the influence factors of the radiopharmaceutical vials dimensions used for activimeter calibration at IPEN.

This paper presents the establishment of a quality control program and correction factors for the geometry of the vials used for distribution of radiopharmaceutical and activimeters calibration. The radiopharmaceutical produced by IPEN 67Ga, 131I, 201Tl and 99mTc had been tested using two different vials. Results show a maximum variation of 22% for 201Tl… CONTINUE READING