Contact finite element stress analysis of porous ingrowth acetabular cup implantation, ingrowth, and loosening.

@article{Rapperport1987ContactFE,
  title={Contact finite element stress analysis of porous ingrowth acetabular cup implantation, ingrowth, and loosening.},
  author={D J Rapperport and Dennis R. Carter and David J. Schurman},
  journal={Journal of orthopaedic research : official publication of the Orthopaedic Research Society},
  year={1987},
  volume={5 4},
  pages={
          548-61
        }
}
Two-dimensional linear and contact finite element analyses were conducted of total hip arthroplasty using metal-backed, porous ingrowth acetabular components. The stress transmission characteristics from the component to the surrounding bone were given special attention. Resultant loads of 20 and 40 degrees medial of vertical were studied, and the influence of adding a metal flange to the rim of the cup was evaluated. The results indicated that when a conventional metal-backed component… CONTINUE READING