Consequences of random and systematic reconstruction uncertainties in 3D image based brachytherapy in cervical cancer.

@article{Tanderup2008ConsequencesOR,
  title={Consequences of random and systematic reconstruction uncertainties in 3D image based brachytherapy in cervical cancer.},
  author={Kari Tanderup and Taran Paulsen Hellebust and Stefan Lang and J{\o}rgen Granfeldt and Richard Poetter and Jacob Christian Lindegaard and Christian Kirisits},
  journal={Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology},
  year={2008},
  volume={89 2},
  pages={
          156-63
        }
}
BACKGROUND AND PURPOSE The purpose of this study was to evaluate the impact of random and systematic applicator reconstruction uncertainties on DVH parameters in brachytherapy for cervical cancer. MATERIAL AND METHODS Dose plans were analysed for 20 cervical cancer patients with MRI based brachytherapy. Uncertainty of applicator reconstruction was modelled by translating and rotating the applicator. Changes in DVH parameters per mm of applicator displacement were evaluated for GTV, CTV… CONTINUE READING